RD SET™
Adapter Cable Series
Reusable
INDICATIONS
The RD SET™ Adapter Cable Series have the same indications for use as the applicable sensors. Please see the compatible
sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
The RD SET Adapter cables allow Masimo RD SET sensors to be used on PC or M-LNC and RC patient cables on Masimo®
SET® or Masimo rainbow® SET devices. The RD SET Adapter cables allow Masimo® SET® LNOP®, LNCS® or M-LNCS™
sensors to be used with RD SET series patient cables. RD EXT cables extend the length of RD SET and RD rainbow SET
series patient cables.
RD SET Adapter Cable Series consist of the following:
• RD to PC Adapter Cable
• RD to M-LNC Adapter Cable
The RD SET Adapter Cable Series have been verified using Masimo SET or Masimo rainbow SET technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET or Masimo rainbow
SET oximetry or licensed to use Masimo sensors.
NOTE: When using the RD SET EXT cable with a RD rainbow SET series patient cable, the maximum length for the
sensor, cable and extension cable combination is 15 feet.
WARNINGS
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable
and discard if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable
and sensor before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings,
inaccurate results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or
disconnecting either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize
the cable or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may
damage the electrical components, potentially leading to patient harm.
INSTRUCTIONS
A) Attaching the RD SET Adapter Cable to a Patient Cable
RD to PC Adapter Cable
1. Refer to Fig. 1. Orient the connecting tab so that the "shiny" contacts are facing up and mate to the patient cable.
2. Insert the connector into the patient cable connector until there is a tactile or audible click of connection.
3. Gently tug on the connectors to ensure a positive contact.
RD to M-LNC Adapter Cable
1. Refer to Fig 2. Properly orient and insert the connector completely into the patient cable connector.
2. Refer to Fig. 3. Completely close the protective latch.
PC to RD Adapter Cable / LNCS to RD Adapter Cable / M-LNCS to RD Adapter Cable / RD SET EXT-4:
1. Refer to Fig. 4. Properly orient the connector and insert the connector completely into the patient cable connector.
2. Refer to Fig. 5. Close the protective latch completely.

DIRECTIONS FOR USE

Not made with natural rubber latex
LATEX
PCX-2108A
02/13
• LNOP to RD Adapter Cables
• M-LNCS to RD Adapter Cables
• LNCS to RD Adapter Cables
4
Non-Sterile
• RD SET EXT-4
9147F-eIFU-0616
en