Table of Contents
1 Specifications
on page 3
2 General information
on page 3
3 Installation
on page 5
4 User interface
on page 6
5 Startup
on page 7
Section 1 Specifications
Specifications are subject to change without notice.
Table 1 General Specifications
Specification
Details
Operating
5 to 40 ºC (41 to 104 ºF)
temperature
Altitude limit
2000 m (6500 ft)
Pollution degree
2
Installation
II
category
Storage/operating
Maximum relative humidity is 80% for temperatures up to
humidity
31 ºC (87.8 ºF), decreases linearly to 50% relative
humidity at 40 ºC (104 ºF)
Location
Laboratory / Indoor
Protection class
2
Range
Selectable, 0 to 35, 0 to 70, 0 to 350, 0 to 700 mg/L
Dimensions
28.9 x 26 x 9.8 cm (11.375 x 10.25 x 3.875 in.)
External power
Input: 100 to 240 VAC, 50/60 Hz, 1.5 A; Output: 24 VDC,
supply
2.7 A, UL CSA and TUV approved.
Capacity
Six 492 mL bottles
6 Standard operation
on page 7
7 Maintenance
on page 9
8 Troubleshooting
on page 10
9 Replacement parts and accessories
on page 11
Table 1 General Specifications (continued)
Specification
Details
Shipping weight
4 kg (8.8 lb)
Warranty
1 year (EU 2 years)
Table 2 Method performance specifications
Specification
Details
Precision
Parameters:
• Standard: 150 mg/L each of glucose and glutamic acid
• Number of samples: 44
• Number of analysts: 1
• Number of BodTrak II instruments: 6
Results:
• Mean of 235 mg/L BOD
• Distribution: 11 mg/L or range of 224 to 246 mg/L BOD
• 95% confidence limit
Drift
Less than 3 mg/L BOD in 5 days
Resolution
1 mg/L BOD
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer's
website.
2.1 Expanded manual version
For additional information, refer to the CD for an expanded version of
this manual.
English 3