TA368 Current Probe User's Guide
3

SPECIFICATIONS

Measuring range
Output sensitivity
(AC coupled)
Accuracy (45–65 Hz)
Minimum load resistance
Frequency range
Phase (45–65 Hz)
Position sensitivity
External field
Noise
Temperature coefficient
Power supply
General characteristics
Probe length
Probe diameter
Output cable
Material
Degree of Protection
Operating temperature
Storage temperature
Operating humidity
Safety standards
EMC
Environmental standards
6
2000 A AC peak
0.7 mV per A
±1% of reading ± 1 A
100 kΩ for specified accuracy
10 Hz to 20 kHz (–1 dB)
< ±1°
±2% of reading
±0.2% of range with cable > 100 mm from the probe
0.6 A typical
±0.1% / °C
5 V nominal via PicoScope 4444
610 mm (24"), double-insulated
9.9 mm (0.39")
2 m (78.7")
Alcryn 2070 NC, LATI LATENE 7H2W V0
IP65
–20 °C to +65 °C (–4 °F to +149 °F)
–40 °C to +75 °C (–40 °F to +167 °F)
15% to 85% (non-condensing)
EN 61010-1
EN 61010-2-032
1000 V RMS, Category III, 600 V RMS, Category IV
Rated for continuous use
EN 61326-2-2
RoHS and WEEE compliant
Pollution Degree 2