Description of your controller's symbols:
There are a number of technical markings on your controller.
These can be explained as follows:
Inside the battery compartment '+' indicates positive
polarity and '-' indicates negative polarity. The batteries
should be positioned as indicated by three embossed
outlines in the base of the compartment.
The controller and belt are made in China for Bio-Medical
Research Ltd., Parkmore Business Park West, Galway,
Ireland.
The controller requires three 1.5 AAA volt DC batteries. DC
is indicated by the symbol:
This symbol means "Attention, consult the
!
accompanying documents".
This symbol means type BF equipment (GelPads).
SN stands for "serial number". On the label, on the back
of the controller is the serial number specific to this
product. The letter preceding the number indicates the
year of manufacture, where 'U' denotes 2015, 'V'
denotes 2016 etc.
The batch number of the belt is represented on the belt
packaging, by the number corresponding with the LOT
symbol.
www.slendertone.com
English
P
ROGRAM PARAMETERS
• Output Current
• Pulse Rate
• Pulse Duration (Width)
• Mode # 1
• Mode # 2
• Mode # 3
• Mode # 4
• Mode # 5
• Mode # 6
• Mode # 7
• Mode # 8
• Mode # 9
• Mode # 10
• Timer
• Compliance
• Waveform
• Maximum Charge Per Pulse
• Power Source
• Tolerances
0-70 milliamps, adjustable (up to 1K ohms load)
50, 55, 60, 70, 75 and 99 pulses per second (PPS)
Positive pulse width 200–350 microseconds / Negative pulse width
200–350 microseconds / Inter-phase interval 100 ± 40 microseconds
Pulse rate 70 PPS / On time = 2.5 seconds / Off time = 2.5 seconds
Ramp Up and Down = 2 seconds / Timer = 20 minutes
Pulse rate 60 PPS / On time = 3.5 seconds / Off time = 3.5 seconds
Ramp Up and Down = 2 seconds / Timer = 25 minutes
Pulse rate 50 PPS / On time = 4.5 seconds / Off time = 4.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
Pulse rate 50 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 2 seconds / Timer = 40 minutes
Pulse rate 55 PPS / On time = 6.5 seconds / Off time = 6.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
Pulse rate 55 PPS / On time = 7.5 seconds / Off time = 7.5 seconds
Ramp Up and Down = 1-2 second / Timer = 30 minutes
Pulse rate 55-75 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 0.4-2 seconds / Timer = 35 minutes
Pulse rate 55-75 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 0.4-2 seconds / Timer = 40 minutes
Pulse rate 70 PPS / On time = 3 seconds / Off time = 3 seconds
Ramp Up and Down = 2 seconds / Timer = N/A
Pulse rate 99 PPS / On time = 2 seconds / Off time = 2 seconds
Ramp Up and Down = 2 seconds / Timer = N/A
20, 25, 30, 35 and 40 minutes depending on the mode.
Complies with 60601-1
Symmetrical biphasic with zero net DC
15 microcoulombs under normal conditions
3 x 1.5V AAA Batteries
All electrical specifications are ± 10% into real 500-ohm load
15