7. Disinfection procedure of the camera head
Pre-disinfection cleaning
Immersion in an enzymatic cleaning
solution (Aniosyme DD1, Hexanios
G+R or surface cleaning (Anios wipe))
Any other method of disinfection is prohibited. Damage caused by these other methods cannot be
borne by the manufacturer.
The use of a sterile protective cover on the camera is recommended regardless of the procedure fol-
lowed in stetilization service and for the duration of the surgical procedure.
The camera is not autoclavable.
The camera is not compatible with automatic washers/disinfectors.
Existing alkaline solutions for pre-disinfection of certain medical devices are FORBIDDEN for pre-di-
sinfection of our cameras.
Parts that have been in contact with the disinfectant must be rinsed thoroughly.
Use non-woven compresses to dry the optics so as not to scratch them.
The procedures described in this chapter are provided as advice, not in any way to replace official
recommendations or guidelines.
Decontamination being linked to the selected products, methods and/or tools, remains the sole res-
ponsibility of the personnel concerned.
8. Failure detection
8.1.
The light indicator of the front panel does not illuminate when power is turned on
- Check that the power cord is connected to the network and the device, and that the power switch on the
rear panel of the unit is in the « I » position.
- Check fuses for proper condition (use T2A - 250V - UL/CSA fuses only).
8.2. The LED source lights up but the luminous flux is insufficient
Check if the light source power is set to Pmax (100% of the power) thanks to the button of the camera head. If
the problem remains even with Pmax value, check if the light cable is properly plugged. Check eventually the
condition of your light cable and optics.
8.3.
There is no more light but the fans continue to work
Check that the light cable is properly connected. If this is the case, wait a few minutes: the light source is equip-
ped with a safety device that cuts off the power supply to the LED if the temperature inside the device is too high.
Once the temperature has dropped, the source can be used again.
As a reminder, there must be enough space around the unit to ensure sufficient cooling (15cm all around) and
nothing must obstruct the fan grids below, on the back and on the left side of the device.
D900 700 053 B
Instructions for use: iCare imaging station
Immersion in a glutaraldehyde so-
lution (Anios-Steranios Laboratory
Rev 09/2018
Disinfection
2%)
Sterilization
Sterilization with ethylene oxide
delmont
imaging
10