6. OPERATION
6.1 POWERING UP THE LIGHT SOURCE
To operate the light source: Attach the light source to the instrument or headlamp.
Turn on the rear power switch. Turn on the lamp switch. The fan will start. The indicator on
lamp switch will light. The lamp will ignite.
6.2 LIGHT BRIGHTNESS DISPLAY AND CONTROL
The light brightness adjustment enables the user to obtain a brighter or darker illumination of
the object of observation.
Adjust the light intensity by turning the intensity control knob.
6.3 LAMP LIFE HOURS DISPLAY
The hour meter shows the total elapsed running time of the unit.
7. CLEANING AND DISINFECTION
NOTE
Always disconnect the power cord before cleaning and while unit is
drying if wet-wiped.
NOTE: Always disconnect the power cord before cleaning and while unit is drying if wet-wiped.
Cleaning and disinfection agents should be commercially available cleansers commonly used
for cleaning and/or disinfection of electronic equipment cabinets in hospitals, such as ethyl or
isopropyl alcohols, disinfecting sprays containing quaternary ammonium compounds or
hydrogen peroxide.
Do not use strongly caustic or acidic cleaners such as "Clorox" hypochlorite bleach, ammonia,
muriatic acid or similar products. Do not use acetone, methyl ethyl ketones or halogenated /
chlorinated hydrocarbon solvents, or cleaners containing any of these restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids over the cabinet.
Do not allow liquids to enter the cabinet seams or ventilation openings, or remain within switch
or control device crevices
8. MAINTENANCE, SERVICING AND REPAIR, WARRANTY
Performance of preventive maintenance is not essential. We recommend that once every two
months that your biomedical or other hospital technical people remove the lamp module from
the light source and use compressed air to blow off any accumulated lint. Regular maintenance
can, however, contribute to identifying potential problems before they become serious; thus,
enhancing the instrument's reliability and extending its useful operating life. Maintenance
services can be obtained from your local representative or from the manufacturer.
Defective items of equipment are to be serviced and repaired exclusively by persons
authorized by the manufacturer. All repair work shall employ original manufacturer's parts only.
LIT124SUNOPTIC SURGICAL
(English)
®
Rev. I
Date of Revision: 02/08/16
Page 9 of 66