Indicated by a flashing red droplet and red 'X'. The humidifier will be disabled during therapy.
This indicates a corruption, most likely caused by moisture in the CPAP. Allowing the
system to dry out may correct the issue. To do this, run the CPAP in Dry Mode.
NOTE:
After drying the system, disconnect and reconnect the power source to
ensure the status is reset.
Heater Status Indicators
These appear in lower right portion of the Home Screen (next to the humidification icon)
Flashing Red X – Heater Detached or Damaged
Flashing red X
Contact your supplier for repair or replacement.
If assistance is needed to reset an indicator, contact your medical equipment
provider or Somnetics customer service 877-621-9626 or info@somnetics.com
Combination Indicators
The Water and Heater status symbols are not mutually exclusive. It is possible that both
issues may occur and will be shown in combination in the lower right portion of the LCD
screen. Example: