The SUPERVEGA EVO suction equipment does not need maintenance or lubrication.
It is, however to inspect the unit before each use. With regard to training, given the information contained in the user manual and
since it is easy to understand the said device, it doesn't appear to be necessary. Unpack the instrument and always check integrity
of plastic parts, they might have been damaged during previous use.
Turn switch on and verify that loud noises are not present, these can indicate wrong functioning.
Make sure you do not hear excessively annoying noises that may highlight a malfunction.
Internally, the device (see electrical sheet) is protected by two fuses F1 and F2 (F 15A L 125V), which cannot be reached from the
outside; refer to technical personnel authorised by the manufacturer to replace them.
The device is equipped with a Lithium-Ion Battery, which cannot be reached from the outside. Only refer to GIMA technical service
personnel to replace it. The device should be checked at least once every 12 months by technical service. Every 24 months it is
compulsory to have a safety inspection and technical maintenance performed on the device.
ONLY USE BATTERIES RECOMMENDED BY GIMA. USE OF BATTERIES OTHER THAN THOSE RECOMMENDED
WILL MAKE THE WARRANTY VOID AND NULL.
ONLY QUALIFIED PERSONNEL ARE AUTHORISED TO REPLACE THE INTERNAL BATTERY. ANY OPERATION
PERFORMED BY NON-TRAINED PERSONNEL MAY CAUSE DANGER (E.G. EXCESSIVE TEMPERATURE)
Fault type
1. Fixed Red Back-Light
2. No LED is On and the Back-
Light failed to switch on
3. No aspiration
4. No aspiration
5. The float doesn't close
6. The float doesn't close
7. Low suction
8. No aspiration due to flow
leakage of mucus
9. The Vacuum power on the
patient side is either very low
or absent
10. Noisy
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 –
9 - 10
Lithium-Ion Battery Charging Cycles: The Lithium-Ion battery contained inside the device is guaranteed for a number greater
than 300 charging cycles. In the vicinity of 300 charging cycles you can ask for verification of operating status to the manufacturer
or to require replacement of the battery pack in such a way that you always have the component in perfect condition.
If the overfill security system it's activated, don't proceede with the liquid aspiration. If the overfill security system doesn't work
there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device, in this case return the
device to GIMA technical service.
GIMA S.p.A. will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.

PERIODICAL MAINTENANCE CHECKS

Cause
Low battery
Locking device
Jar Cap badly screwed down
Lid seal not in its seat
If the cap has been washed,
ensure that the float is not
partially detached
The float it's covered by dirty
material
Foam inside the jar
Filter blocked
 Protection filter blocked or
damaged
 Connection tubes blocked,
kinked or disconnected
 Shut-off valve blocked or
damaged
 Pump motor damaged
Technical internal problem
None of the remedies has
achieved the desired results
If it is intended to be used in ambulances, position the
device on the support bracket and leave the battery charge
until the Fixed White Back-Light gives the signal (TAB. I).
Internal technical problem. Contact technical assistance.
Unscrewed the cap, then rescrew it correctly
Unscrew the cap and insert the seal properly in its seat
Insert the float into it's place
Unscrewed the cap, leave the and put in on autoclave
Fill the jar to 1/3 full of ordinary water
 Replace the filter
 Replace or reconnect the tubes, check the jar
connections
 Empty the jar, or disconnect the tube from the jar and
unblock the shut-off valve. The unit twill only work in
the upright position
 Contact the technical service
Contcat the technical service
Contact the seller or GIMA After-sales Assistance
23
Solution
Replace filter
Service
GB