Two Yellow Collecting Waves
Description:
Action:
The Communicator started but was
not able to complete the interrogation
within the time allowed.
• Ensure the Communicator is
optimally placed as described
in "Where to Place Your
Communicator" on page 13.
• Face the Communicator. Sit directly
in front of the Communicator. Make
sure you are within 3 feet (1 meter)
of the Communicator.
• Remain still until the interrogation is
complete. Do not move away from
the Communicator.
• Move any wireless electronic
products (such as cordless or
cellular phones or baby monitors)
that are within 3 feet (1 meter) of the
Communicator.
To verify that troubleshooting was a
success:
• Press the Heart button to start
another interrogation. If the
interrogation was a success, all
three Collecting Waves will light
green for 2 minutes.