Sensor Calibration

(PATH: Device setup / Sensor Calibration)
h
Note: For best process calibration results observe the following points. Take grab sample as
close as possible to the measurement point of the sensor. Measure the sample at process
temperature.
Via configuration tool, asset-management-tool or HART handheld terminal you can calibrate
the sensor with the "Process", "1-Point" or "2-Point" calibration method. For the calibration
methods see Operating Instructions of the M100 transmitter. Via iSense software you can
calibrate the sensor with the "1-Point" or "2-Point" calibrate method. See Operating Instructions
of the iSense software.
As soon as the calibration is in progress no other calibration can be started.

Sensor Calibration Menu

For the calibration methods see "Menu Structure" chapter on Page 37.
After every successful calibration the following options are available:
• Adjust
Calibration values are adopted and used for the measurement. Additionally, the calibration
values are stored in the calibration history.
• Calibrate
Calibration values are stored in the calibration history for documentation, but not be used
for the measurement. The calibration values from the last valid adjustment are further used
for the measurement.
• Abort
Calibration values are discarded.