RS-232 INTERFACE
With a standard build-in RS-232 interface, the balance is able to be
connected to a printer or computer or other peripheral device for outputting
weighing data.
Transmission Format
Transmission form: asynchronous transmission
baud rate: 1200, 2400, 4800, 9600
Start bit: 1
Stop bit: 1
Data bit: 8
Parity: none
Code: ASCII
Data output format:
ST, +100.0000¡¡g <CR> <LF> <LF> or
US, +100.0000¡¡g <CR> <LF> <LF>
• ST=STABLE (53H, 54H)
• US=UNSTABLE (55H, 53H)
• ¡=SPACE (20H)
• <CR>=CARRIAGE RETURN (0DH)
• <LF>=LINE FEED (0AH)
Data output
Auto Output Mode
There are three modes:
16