In addition, please compare the accessories you receive with the standard accessories list below. If
an item is missing, please notify BUFFALO FILTER Customer Service.
Standard Accessories:
• Operator's Manual
• Power Cord
• Pneumatic Footswitch
Please contact BUFFALO FILTER Customer Service to purchase the following accessories:
• Replacement Filters
• EZ Link™ Remote Activator
• Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories
1.3 Operational Information
The operational information contained in this section is intended for the customer review of
regulatory issues. The information pertains to the use of the products both domestically and
internationally:
1. The BUFFALO FILTER Turbo Auto-Sense™ Smoke Evacuation System(s) complies with
IEC60601.1 electrical specifications in the following systems:
100/120 VAC
50/60 Hz,
2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I
3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part
4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1
5. Method of sterilization or disinfection recommended by BUFFALO FILTER (UL 60601-1,
Clause 5.4):
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution
or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
6. Degree of safety of application in the presence of flammable anesthetic mixture with air or
with oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable
7. Mode of operation (UL 60601-1, Clause 5.6): Continuous
8. Upon request, BUFFALO FILTER will provide the following:
Service and Repair Instructions, including Circuit Diagrams and Parts List
9. The fuses in the power entry module are to be serviced by an authorized BUFFALO FILTER
technician as follows:
100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo)
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo)
220/240 VAC 50/60 Hz