Look inside the transmitter's connector opening to make sure that the connector pins are
not damaged or corroded. If the connector pins are damaged or corroded, the transmitter
cannot communicate with the charger or pump. Contact your local help line or
representative for assistance. It may be time to replace your transmitter.
Also look for moisture inside the connector opening. If you see any moisture, allow the
transmitter to dry for at least one hour. Moisture inside the connector opening could
cause the transmitter to not work properly, and could cause corrosion and damage over
time.
Connecting the tester for testing or cleaning
To connect the tester:
1
Hold the transmitter and the tester as shown. Line up the flat
side of the tester with the flat side of the transmitter.
2
Push the tester into the transmitter until the flexible side arms of
the tester click into the notches on both sides of the transmitter.
Within 20 seconds the green light on the transmitter will flash
for about 10 seconds when properly connected.
3
To test the transmitter, check the sensor icon on the pump or
monitor to ensure that the transmitter is sending a signal (see
the user guide for your device).
4
To clean the transmitter, see Cleaning/disinfecting the transmitter, on page 9.
5
After testing or cleaning, disconnect the tester from the transmitter.
Disconnecting the tester
To disconnect the tester:
1
Hold the transmitter body as shown and pinch the side arms of
the tester.
2
With the tester arms pinched, gently pull the transmitter away
from the tester.
Note: To save transmitter battery life, do NOT leave the tester
connected after cleaning or testing.