Cleaning the charger
The charger cannot be disinfected. This procedure is for general cleaning as required,
based on physical appearance.
Caution:
The charger is NOT waterproof. Do NOT immerse in water or any other
cleaning agent.
Caution:
Dispose the charger according to the local regulations for battery
disposal (non-incineration).
Warning:
For multiple-patient use, always clean and disinfect the transmitter after
removing it from the patient and before attaching it to the charger. If
blood comes in contact with any surface areas of the charger, the
contaminated device must be discarded. The charger contains a battery
which may explode upon incineration.
To clean the charger:
1
Wash your hands thoroughly.
2
Use a damp cloth with mild cleaning solution, such as a dishwashing detergent, to
clean any dirt or foreign material from the outside of the charger. Never use organic
solvents, such as paint thinner or acetone, to clean the charger.
3
Place the charger on a clean, dry cloth and air dry for 2–3 minutes.
Troubleshooting
Question: Why do I see quick flashing red lights on the charger?
Answer: The transmitter battery is very low. Leave the transmitter on the charger for
8 hours to completely recharge. If the red light is still flashing after 8 hours, leave your
transmitter on the charger for 24 hours. If the red light is still flashing after it has been
charging continuously for 24 hours, contact your local help line or representative for
assistance. It may be time to replace your transmitter.