Descargar Imprimir esta página

Agilent Technologies Turbo-V 81-M Serie Manual De Instrucciones página 219

Publicidad

1
Figure 25
2
Cooling may be carried out either through an open circuit with
eventual discharge of the water, or using a closed circuit cooling
system.
The water temperature must be between +15 °C and +35 °C, with an
inlet pressure between 2 and 4 bar.
The water electrical conductance must be ≤ 500 µs/cm. When the conductance
NOTE
is higher, in closed water circuit, the use of up to 20 % of Ethyl-Glycole is
suggested.
Turbo-V 81-M User Manual / 87-900-983-01 (F)
Connect the plate to the pump bottom with four screws M3x20.
Assemble the metal or the plastic kit as shown.
Technical Information
Water Cooling Kit Installation
17
219/237

Publicidad

loading

Este manual también es adecuado para:

969-8901969-8902969-8903969-8904