CHANNEL A, ITS INTENSITY WILL DROP TO "0" AND THEN, ONCE THE 20
MINUTES ARE UP FOR CHANNEL B, THIS WILL GO TO "0" AND THE EQUIPMENT
WILL AUTOMATICALLY SWITCH OFF. IF THE SYSTEM IS SWITCHED ON, BUT IS
NOT ACTIVATED, THE APPLIANCE WILL SWITCH OFF AFTER 4 MINUTES.
INTENSITY CONTROL
THE INTENSITY CONTROL SCALE IS SPLIT INTO THE FOLLOWING 18 LEVELS:
1 flashing = intensity level 0,5
1 steady = intensity level 1
2 flashing = intensity level 1,5
2 steady = intensity level 2
3 flashing = intensity level 2,5
3 steady = intensity level 3
4 flashing = intensity level 3,5
4 steady = intensity level 4
5 flashing = intensity level 4,5
5 steady = intensity level 5
6 flashing = intensity level 5,5
6 steady = intensity level 6
7 flashing = intensity level 6,5
7 steady = intensity level 7
8 flashing = intensity level 7,5
8 steady = intensity level 8
9 flashing = intensity level 8,5
9 steady = intensity level 9 (maximum intensity)
Do not use T APPING (I) or T APPING (II) on one channel at the same time as
KNEADING, PRESS or VIBRATION on the other channel.
ERROR FUNCTION (DETACHED ELECTRODE): IF AN ELECTRODE IS DET ACHED,
THE APPLIANCE REGISTERS AN ERROR. ONLY "E E" (FLASHING) WILL APPEAR
ON THE DISPLA Y TO INDICA TE AN ANOMALY AND ALL OUTPUT SUPPL Y WILL
BE CUT. TO RESET THE SYSTEM, SWITCH IT OFF , CONNECT THE ELECTRODE
AND SWITCH BACK ON. OTHERWISE, THE SYSTEM WILL AUTOMA TICALLY
SWITCH OFF AFTER 4 MINUTES. N.B. WHEN USING THE DOUBLE CABLE FOR 8
ELECTRODES, THIS FUNCTION WILL ONLY WORK IF BOTH THE 2 ELECTRODES
OF THE SAME POLARITY ARE DET ACHED (2 BLACK ONES OR 2 RED ONES).
THIS FUNCTION IS INFLUENCED BY THE DIFFERENCE IN RESIST ANCE IN EACH
INDIVIDUAL AND SO THE LOW INTENSITY DETECTION TIME MA Y BE INACTIVE.
DEMO FUNCTION: THIS FUNCTION IS FOR USING THE APPLIANCE WITHOUT
ANY CURRENT AND WITHOUT ANY ERROR SIGNALS. TO USE THIS FUNCTION,
SWITCH THE APPLIANCE OFF AND THEN PRESS KEY 4 FOR 5 SECONDS. YOU
ARE IN DEMO MODE WHEN INTENSITY = 8 8 AND CHANNELS A-B APPEAR
WITH THE TWO FLASHING ARROWS POINTING TO T APPING 1. YOU CAN VIEW
ANY PARAMETER FOR A CHOSEN PROGRAM WITHOUT ACTIVATING THE TIMER
FUNCTION OR THE OUTPUT SIGNAL. NB: DEMO FUNCTIONS DO NOT AL WAYS
MATCH THE REAL ONES. TO QUIT DEMO MODE, SWITCH THE SYSTEM OFF
AND THEN SWITCH BACK ON.
RESET FUNCTION: IN CASE OF MALFUNCTIONING RESET THE TESMED
AFTER ANY MALFUNCTION BY SIMPL Y REMOVING AND REINSER TING THE
BATTERIES.