STEP 3. Obtain Control Solution
Shake the control solution vial well. Remove the
cap from the control solution bottle. Place cap
on flat surface. Squeeze the vial, discard the first
drop, and wipe off the dispenser tip to prevent
contamination. Squeeze the vial again to produce
another drop and place this drop on the top of
cap.
STEP 4. Apply Control Solution
While holding the meter, move the absorbent hole
of the test strip to touch the drop of control solu-
tion. Then the drop will be automatically drawn into
the test strip. Make sure the confirmation window
fills completely. The meter begins to count down.
To avoid contaminating the control solution with
the contents of the test strip, you have to place
a drop of control solution on a clean surface. Do
not directly apply control solution to a strip.
STEP 5. Read and Compare the Result
After counting down to 0, the test result of control
solution is shown on the screen. Compare this
result with the range printed on the test strip vial. It
should fall within this range.
Out-of-range results
If test results fall outside the range printed on the test strip vial, check "Operational
Problem" section in troubleshooting guide and repeat the test. If you continue to
get out-of-range results, it means that the system may not be working properly.
Do NOT test your blood. Contact the Customer Care Line at 1-866-469-2632.
24
Normal
109~165