TROUBLESHOOTING

If any abnormality arises during use, please check the following points.
SYMPTOMS
Unit does not turn
on when the START/
STOP button is
pushed.
EE measurement error
symbol shown on
display or the blood
pressure value is
displayed excessively
low (or high).
E1 error symbol shown
on display
E2 error symbol shown
on display
E3 error symbol shown
on display
EP error symbol
shown on display
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.
POSSIBLE
CAUSES
Batteries have
run down.
Battery polarities
have been positioned
incorrectly.
Cuff has been placed
incorrectly.
Did you talk or move
during measurement?
Shaking of the arm
with the cuff on.
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Inflation pressure
exceeding 300 mmHg.
Error determining
measurement data.
System error.
30
CORRECTION
Replace them with
four new "AA"
alkaline batteries.
Re-insert the batteries
in the correct
positions.
Wrap the cuff properly
so that it is positioned
correctly.
Keep arm steady
during measurement.
Measure again.
Check cuff connection.
Measure again.
Switch the unit off,
then measure again.
Measure again.
Measure again.