Megger DET2/2 Guia Del Usuario página 31

Ocultar thumbs Ver también para DET2/2:
Tabla de contenido

Publicidad

Display:
Instrument Protection:
Temperature Effect:
Temperature Range:
Humidity:
Flash Test:
Voltage Withstand:
Compliance with Standards:
Power Supply:
Battery life:
Battery charging time:
Charging supply required:
Power consumption:
Alpha numeric LCD (130 mm x 35 mm) giving test information and a large
1
(20 mm) 3
digitLCD, maximum reading 1999
2
Meets the general requirements of IP54
<±0,1%/°C over the temperature range -10°C to +40°C
Operating:
-10°C to +40 °C
Storage:
-20°C to +60°C
Operating:
90% RH max. at 40°C
3 kV a.c.
In the event of a system fault the instrument will withstand 240 V a.c.
applied between any two terminals.
BS 7430 (1992)
VDE 0413 Part 7 (1982)
(i) Internal rechargeable sealed lead acid cells 12 V nominal, 2,6 Ah
capacity. Battery voltage range over which basic accuracy is
maintained, 11,0 V to 13,5 V.
Typically 5 Hours continuous use
6 hours max. (from completely exhausted).
100 V to 130 V or 200 V to 260 V a.c. 50 Hz/60 Hz.
25 VA
BS7671 (1992)
NFC 15 -100
IEC364
31