Error message
Displayed Error
Definition
FLASH FAILURE
The external serial flash
memory Read/Write fails
Event Log
Refer to
Table 3
for diagnostic device information.
Table 3 Event log
Event
0: Power Up Event
1: Device Comm Loss
2: Device Comm Restore
3: Software Restart Event
Classified errors and classified status
Table
4,
Table
5,
Table
6,
Table 7
register and the classified status 1-4 register flags for main
measurements. All the sensors and analyzers supply this signal quality
information at the same register address.
Table 4 Classified errors – register 49930
Bit
Error
0
Measurement Calibration
Error
1
Electronic Adjustment Error
2
Cleaning Error
10 English
Resolution
Contact technical service
Description
Logs the power up time
Reports the loss of communications
with a device. (Data: Device Index)
Reports the restored communications
with a device (Data: Device Index)
Reports a software restart.
and
Table 8
show the classified error
Note
An error has occurred during the last
calibration.
An error has occurred during the last
electronic calibration.
The last cleaning cycle failed.
Table 4 Classified errors – register 49930 (continued)
Bit
Error
3
Measuring Module Error
4
System Re-initialization
Error
5
Hardware Error
6
Internal Communication
Error
7
Humidity Error
8
Temperature Error
9
Reserved for future use
10
Sample Warning
11
Questionable Calibration
Warning
12
Questionable Measurement
Warning
13
Safety Warning
14
Reagent Warning
15
Maintenance Required
Warning
Note
A failure has been detected in the
Measuring Module.
Some settings have been detected
inconsistent and set to factory defaults.
Any hardware error in general has been
detected.
A communication failure within the device
has been detected.
Excessive humidity has been detected in
this device.
Temperature within the device exceeds
specified limit.
Fixed at 0
Some action is required with the sample
system.
The last calibration was of questionable
accuracy.
One or more of the measurements of the
device are of questionable accuracy
(Quality bad or out of range).
A condition has been detected which may
result in a safety hazard.
Some action is required with the reagent
system.
Maintenance is required on this device.