Table 5 Classified status 1 – register 49931
Bit
Error
0
Calibration in progress
1
Cleaning in progress
2
Service/Maintenance menu
3
Common error
4
Measurement 0 Quality Bad
5
Measurement 0 Low Limit
6
Measurement 0 High Limit
7
Measurement 1 Quality Bad
8
Measurement 1 Low Limit
9
Measurement 1 High Limit
10
Measurement 2 Quality Bad
11
Measurement 2 Low Limit
12
Measurement 2 High Limit
13
Measurement 3 Quality Bad
14
Measurement 3 Low Limit
15
Measurement 3 High Limit
Note
The device has been placed in a
calibration mode. The measurements may
not be valid.
The device has been placed in a cleaning
mode. The measurements may not be
valid.
The device has been placed in a service or
maintenance mode in which the
measurements may not be valid.
Device recognised an error; see Error
Register for Error Class.
Precision of measurement is out of
specified limits.
Measurement is below measurement
range.
Measurement is above measurement
range.
Table 6 Classified status 2 – register 49932
Bit
Error
0
Reserved for future use
1
2
3
4
Measurement 4 Quality Bad
5
Measurement 4 Low Limit
6
Measurement 4 High Limit
7
Measurement 5 Quality Bad
8
Measurement 5 Low Limit
9
Measurement 5 High Limit
10
Measurement 6 Quality Bad
11
Measurement 6 Low Limit
12
Measurement 6 High Limit
13
Measurement 7 Quality Bad
14
Measurement 7 Low Limit
15
Measurement 7 High Limit
Table 7 Classified status 3 – register 49933
Bit
Error
0
Reserved for future use
1
2
Note
Fixed at 0
Precision of measurement is out of
specified limits.
Measurement is below measurement
range.
Measurement is above measurement
range.
Note
Fixed at 0
English 11