guiDancE & manufacturEr'S
DEclaration
Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and this device. This device and system are NOT
Life-Supporting
This device is intended for use in the electromagnetic environment in which radiated
disturbances are controlled. The customer or user of this device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF Communications Equipment and the device as recommended below,
according to the maximum output power of the communications equipment.
Recommended Separation Distances for the device (meters)
Maximum
Output
150 kHz to 80 MHz
Power
(Watts)
d=(1.1667)
0.01
0.11667
0.1
0.36894
1
1.1667
10
3.6894
100
11.667
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
80 to 800MHz
√
P
d=(1.1667)
0.11667
0.36894
1.1667
3.6894
11.667
800 MHz to 2.5 GHz
√
P
d=(2.3333)
0.23333
0.73785
2.3333
7.3785
23.333
EN
√
P
SE-DV56
31