3B SCIENTIFIC PHYSICS 1000642 Instrucciones De Uso página 9

Juego complementario rmn
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
Fig. 5 Pulling the magnets onto the assembly discs
Fig. 6 Basic unit completed with permanent magnets
and coils
7.1.1 Removing magnet unit
Take the sample out of the sample chamber
in the basic unit.
Disconnect the cables between the coils and
the control console.
Loosen the knurled screws.
Turn the basic unit so that the magnet unit is
pointing forwards.
Lift up the yoke till it is on top of the sample.
Hold the yoke in place with your thumbs and
use your fingers to pull the magnet unit to-
wards the front till the yoke is between the
two magnets. Then take the entire unit out of
the basic unit (see Fig. 7).
Take the discs out of the sample chamber.
Fig. 7 Removing the magnet unit from the basic unit
7.2 Connection to control console
Insert the probe into the sample chamber in
such a way that it rests on the housing (see
Fig. 8).
Connect the cable for the probe to the
"Probe In" socket of the control console, tak-
ing note of the socket's slot.
Connect the coils to the "Coil" sockets on
the back of the control console.
Connect the control console to its power
supply via the "12 VAC/1A" socket.
Fig. 8 Basic unit with probe
7.3 Calibration and settings
7.3.1 Using an oscilloscope
Connect the "SIGNAL OUT" socket of the
control console to channel 1 of the oscillo-
scope and the "FIELD OUT" output to chan-
nel 2 (see Fig. 12).
Set the oscilloscope as follows:
Channel 1: 0.5 V DC
Channel 2: 0.5 V DC
Time base: 5 ms
Trigger from a falling edge on channel 1.
7.3.2 Using 3B NETlog™
Connect the "SIGNAL OUT" socket of the
control console to input
log™ unit and "FIELD OUT" to the input
Connect the 3B NETlog™ unit to a computer
and run the 3BNETlab
Create a new data record from the "Meas-
urement lab" menu and define the following
parameters:
Input A: Field, Input mode VDC, Input range 2 V
Input B: Signal, Input mode VDC, Input
range 2 V
Measurement interval: 500 µs (2 kHz)
Set to trigger from Input A with a falling edge
and a positive trigger point at about 10 to 20%.
Activate the "Oscilloscope" button and start
measuring.
The oscilloscope window will open.
3
IN
on the 3B NET-
U
B
IN
.
U
A
TM
software

Publicidad

Tabla de contenido
loading

Tabla de contenido