physician / clinician. (If you don't feel anything, turn the Knob OFF then
ON again and carefully turn the Control Knob until you feel a tingling or
slight twitch under or around the electrodes.) Always start with the lowest
setting and increase the intensity slowly.
If the stimulation levels are uncomfortable or become uncomfortable,
reduce the stimulation intensity to a comfortable level; or cease
stimulation and contact your physician.
6.
Setting the Patient Compliance Counter:
a)
To turn on the Patient Compliance Counter: While the unit is ON,
hold down the UP key and press the MODE key simultaneously.
b) To reset the Patient Compliance Counter: While the unit is ON, press
the UP key and the MODE key simultaneously (this will take you into
the Compliance Counter), then push the DOWN key and press the
MODE key simultaneously.
c)
Press the UP and MODE keys simultaneously to return to the
treatment status.
BATTERY INFORMATION
When the LCD (Liquid Crystal Display) low battery mark illuminates, the
battery has become too weak to power the unit, and the existing battery
should be replaced with a new battery. At this point, the unit will turn OFF
until a new battery is inserted.
V NOTICE: GF Health Products, Inc. recommends the use of only a 9V alkaline or
nickel-cadmium rechargeable battery with this device.
Replacing the Battery
When the LCD low battery mark illuminates, and the unit does not remain
illuminated once turned on, the battery should be replaced.
1.
Turn unit OFF .
2.
Remove the front panel cover by pressing on the top of the panel and
pressing down in order to slide the panel down. Continue sliding the
panel downwards until the panel is completely removed from the unit.
This will reveal the battery compartment.
GF-TX5EMS-INS-LAB-RevC14