TECHNICAL SPECIFICATIONS
Channel
Dual, isolated between channels
Pulse Intensity
Adjustable 0 mA – 100 mA peak into 500 ohm load each channel,
constant current
Pulse Width
30μs~260μs, adjustable by 10μs/step
Pulse Rate
1Hz-150Hz,
adjustable
Timer
10, 20, 30, 40, 50, 60 minutes and continuous
Patient Compliance
To show the usage time by user, counted per each 30 minutes treatment
Counter
time
Patient Lock System
To prevent changes in any set parameter by physician / clinician. To
unlock or lock the unit, press the UP and DOWN keys simultaneously for
2-3 seconds
Wave Form
Symmetrical bi-phasic square pulse
LCD
Displays Timer, Function Mode, Pulse Width, Pulse Rate, Cycle ON/
Ramp/OFF Time, Low Battery, Patient Compliance Counter, and Patient
Lock System
Function Mode
C
S
A
Cycle ON Time
(Contraction)
ON Ramp
Cycle OFF Time
(Relaxation)
Voltage
50 volt (peak to peak), at 500 ohm load
Power Source
9 Volt battery
Battery Life
Approximately 70 hours at normal setting
Note: All electrical specifications are ± 10% 500 ohm load.
GF-TX5EMS-INS-LAB-RevC14
1Hz~20Hz adjustable by 1Hz/step
20Hz~150Hz adjustable by 5Hz/step
Constant Mode
Synchronous Mode
Alternating Mode
Adjustable 1-30 second, 1 second/step
Adjustable 1-10 second, 1 second/step
Adjustable 1-30 second, 1 second/step
13