PRODIGY Voice® Blood Glucose Meter
Caution: To avoid contaminating the control
solution with the content of the test strip,
DO NOT DIRECTLY APPLY CONTROL SOLUTION
ONTO THE TEST STRIP. See Figures 20 and 21.
Figure 20
STEP 5 : Check if the Test Result is in Range.
After the meter counts down from 6 to 1, the control
solution test result appears on the display.
Compare the test result with
the range
printed on the test strip vial.
Each vial of Prodigy
No
®
Coding Test Strips may have
a different control solution
range. The result should fall
within the printed range on
the test strip vial.
Owner's Manual
Figure 21
Figure 22
36
PRODIGY Voice® Blood Glucose Meter
See Figure 22.
Questionable Results:
If the test result falls outside the range printed
on the test strip vial, check the "Trouble-Shooting
Guide" located in the "System Troubleshooting"
section and repeat the test.
Results that fall outside of the expected control
ranges may indicate:
• An error in performing the test
• Control solution bottle not shaken well
• Expired or contaminated control solution
• Control solution that is too warm or too cold
• Test strip deterioration
• Meter malfunction
• Test strips expired
Owner's Manual