Recording measured values
Memory function AUTO TRIGGER
The function Auto Trigger documents the process of recording meas-
ured values triggered by a defined start signal (for example, pressure
on channel 2 → 125 bar). In response, a sequence of measured values
are automatically recorded until the previously set measurement time
expires.
Time-dependent functions (for example, making operations or produc-
tion cycles) are measured when recording measured values.
The following parameters must be set:
Menu
TIME UNIT>
DURATION>
PRE
TRIGGER>
TRIGGER>
LEVEL>
TRIG
SLOPE>
AUTO
READY>
112
Setting/
pre-selection
sec, hrs
Number
Number
IN
Number
▲ ▼
YES, NO
Values
Notes
h
Pre-selected time unit
(trigger/pre-trigger)
100
Recording time
1
Pre-trigger time (time
before the trigger signal)
p1
Starting-point measuring
channel
125 bar
Start point value
▲
Ascending or descending
edge
YES
Recording of measured
values is repeated auto-
matically
V2.0/02.11