Cleaning - Roscoe Medical nebGO Manual Del Usuario

Nebulizador portátil ultrasónico
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

• BEFORE connecting power cord, make sure on/off switch is in off position.
• NEVER block the air openings of the main unit or place it where the air openings may be obstructed.
• NEVER allow the power supply cord to be pulled, jerked, strained, twisted or severely bent, espe-
cially at the plug connections. Damage will occur at the high flex point of entry into the appliance,
causing it to rupture and short. Handle the cord carefully for longer life.
• Safety overload: DO NOT exceed 20 minutes of continuous operation.
CLEANING THE NEBULIZER COMPONENTS AND ACCESSORIES
1. If there is any residual moisture (condensation) in the medication cup, wipe it out with gauze or a
lint-free towel.
2. Pour distilled water into the medication cup.
3. Turn on the device to nebulize the distilled water for 1 to 2 minutes to clean the mesh.
4. Power off the device, then remove batteries or disconnect the adapter from the main unit.
5. Remove the medication cup from the main unit.
6. Wash and rinse the medication cup with distilled water.
7. Shake off excess water and allow parts to be fully air dried on a clean, dry towel.
8. Use a clean towel slightly dampened with warm water to clean the outside of the device.
9. Wash the mouthpiece and mask accessories with a mild soap and warm water or soak in 1 part
white vinegar to 3 parts hot tap water for one hour. Rinse thoroughly and allow to air dry completely
before use. Perform disinfection per procedures on page 11.
10. Make sure that all cleaned parts are completely dry before storing or used.
• Keep the battery compartment on the device dry at all times
• DO NOT poke the mesh with finger, cotton swab or any objects.
• DO NOT clean parts in a dishwasher.
• DO NOT use microwave to dry any parts, including device.
• Clean all necessary ports after each use as instructed in this guidebook.
• DO NOT attempt to wash and reuse the air filter. DO NOT use substitute filter material.
Compressor damage may result
• Disconnect the unit from the electrical outlet before cleaning.