Guidance and manufacturer's declaration – emissions
All ME equipment and ME systems
Guidance and manufacturer's declaration – emissions
The Innospire Mini is intended for use in the electromagnetic environment specified below. The customer or user of the
Innospire Mini should ensure that it is used in such an environment.
Emissions test
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonics
Class A
IEC 61000-3-2
Flicker
Complies
IEC61000-3-3
40
Compliance
Electromagnetic environment – guidance
The Innospire Mini uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
The Innospire Mini is suitable for
use in all establishments, including
domestic, and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Guidance and manufacturer's declaration – immunity
All ME equipment and ME systems
Guidance and manufacturer's declaration – immunity
The Innospire Mini is intended for use in the electromagnetic environment specified below. The customer or user of the
Innospire Mini should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
ESD
±6 kV contact
±6 kV contact
IEC 61000-4-2
±8 kV air
±8 kV air
EFT
±2 kV mains
±2 kV mains
IEC 61000-4-4
±2 kV I/Os
±2 kV I/Os
Surge
±1 kV differential
±1 kV differential
IEC 61000-4-5
±2 kV common
±2 kV common
Voltage dips/dropout
95% dip for
100% dip for
0.5 cycle
0.5 cycle
IEC61000-4-11
60% dip for
60% dip for
5 cycles
5 cycles
30% dip for
30% dip for
25 cycles
25 cycles
>95 % dip for
100% dip for
5 seconds
5 seconds
Power frequency
3 A/m
3 A/m
(50/60Hz) magnetic
field
IEC 61000-4-8
ENGLISH
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the r/h should be
at least 30%
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Innospire Mini requires
continued operation during power mains
interruptions, it is recommended that
the Innospire Mini be powered from an
uninterruptible power supply or battery.
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.
41