Charging the battery

1
Charge the battery pack fully prior to first use.
2
Connect the AC or 12V DC car adapter to the power adapter plug-in
on the front side of the battery.
3
Plug the other end of the AC or DC car adapter into the wall outlet
or 12V DC car outlet.
4
The battery indicator light will turn orange.
5
Charge the battery until the battery indicator turns green. It may take up to 6 hours to
fully charge the battery.
6
Unplug the power adapter from both the battery power plug-in and the wall outlet or
12V DC car outlet.

Storing the battery

1
Store the battery fully charged in a cool dry place.
2
Every 1-2 months, fully charge the battery to maintain maximum battery capacity.
3
It may take up to 5 charge/discharge cycles to return the battery to maximum capacity after
prolonged storage.
Battery indicator lights
1
The battery indicator light will turn orange when the
battery is being charged.
2
The battery indicator light will turn green when the
battery is fully charged while still connected to
charging source.
3
The battery indicator light will blink orange when the
battery is almost exhausted/drained.
4
The light is off under the following conditions:
– when the battery has charge and is not connected to
a charging source
– when battery is fully discharged
– when battery enters a fault condition
10
CAUTION: Clean the device, according to instructions, between patients when used on
multiple patients.
Care and cleaning of your InnoSpire Mini
CAUTION: To reduce risk of electric shock, DO NOT DISASSEMBLE and never submerge
or place under running water.
1
Unplug the compressor before cleaning.
2
Use a damp cloth to wipe down the outside of the compressor as required. Do not use
abrasive cleaners.
3
Check the filter periodically and
replace the filter when dirty.
4
To change the filter:
Use a coin to unscrew the filter holder (Note: Turn clockwise to unscrew).
Pull out filter and replace
with new filter.
Indicator light
Screw filter holder back on by turning counter clockwise.
5
After each cleaning, check that the unit is working properly by verifying that air still
comes out of the nozzle when the unit is turned to the "on" position and that the unit's
noise level has not increased. (Increased noise indicates improper filter positioning.)
6
The compressor is oil-free. Do not lubricate.
InnoSpire Mini compressor