Guidance and manufacturer's declaration – immunity
ME equipment and ME systems that are NOT life-supporting
Guidance and manufacturer's declaration – immunity
The Innospire Mini is intended for use in the electromagnetic environment specified below. The customer or user of the
Innospire Mini should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
3Vrms
(V1)=3Vrms
IEC 61000-4-6
150kHz to 80MHz
Radiated RF
3V/m
(E1)=3V/m
IEC 61000-4-3
80MHz to 2.5GHz
42
Electromagnetic environment – guidance
Portable and mobile communications
equipment should be separated from the
Innospire Mini AC by no less than the
distances calculated/listed below:
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
D=(7/E1)(Sqrt P)
where P is the max power in watts and
D is the recommended separation distance
in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance levels
(V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.
Recommended separation distances between portable and mobile RF Communications equipment and the
Innospire Mini
ME equipment and ME systems that are NOT life-supporting
Recommended separations distances for the Innospire Mini
The Innospire Mini is intended for use in the electromagnetic environment in which radiated disturbances are controlled.
The customer or user of the Innospire Mini can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF Communications Equipment and the Innospire Mini as recommended below,
according to the maximum output power of the communications equipment.
Maximum
output power
150kHz to 80MHz
(watts)
D=(3.5/V1)(Sqrt P)
0.01
0.11667
0,.1
0.36894
1
1.1667
10
3.6894
100
11.667
ENGLISH
Separation (m)
80MHz to 800 MHz
800MHz to 2.5GHz
D=(3.5/E1)(Sqrt P)
D=(7/E1)(Sqrt P)
0.11667
0.23333
0.36894
0.73785
1.1667
2.3333
3.6894
7.3785
11.667
23.333
43