electrical inStallation
The sensor to electronics cable is a 3ft, PVC cable for non-cooled housing and 3ft PTFE
cable when the Cooled Housing option is ordered. One end is attached, at the factory,
to the sensor head. The other end has a pair of wires and a shield (ground) wire.
WirinG
The SilverU sensor can be used in 3 different ways as shown below. To use the
4-20mA output, connect the SilverU sensor to the controller/indicator with the 24Vdc
power supply in the loop using the following table and connection diagram as a guide.
The maximum loop impedance is 900 ohms. note: the silveru sensor must be
grounded at only one point, either the cable shield or the sensor housing.
output
Wire no.
4 to 20mA
1
2
3
Alternatively, to use the USB output, connect the SilverU sensor to a PC using the
cable provided. No external power supply is required when using the USB output. The
SilverU sensor can also be used with both the 4-20mA loop and USB connected,
provided that the negative terminal of the 24Vdc power supply being used is not
grounded. note: the silverview software must be installed before connecting the
sensor to a pc.
›
6
Wire colour
function
Red
Power supply + / signal +
Black (Blue)
Power supply - / signal -
Bare
Shield ground
Power Supply
Red
+
-
-
Display/
Controller
+
Black
(Blue)
Power Supply
Red
+
-
-
Display/
Controller
+
Black
(Blue)
USB
PC
N/C
USB
PC
SilVerVieW SoftWare
systeM reQuireMents
•
SilverView software is designed to run under Windows. Suitable versions are
Windows 2000, Windows XP and Windows Vista.
•
The recommended minimum computer specification is:
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
installation
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double click
on Setup.exe)
3. Follow the on-screen instructions
note Do not connect the silveru sensor to the pc before the software is
installed
connectinG the silveru sensor to a pc
1. Connect the circular connector on the USB cable provided to the SilverU sensor
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the SilverView desktop icon
note Do not disconnect the usb cable before closing down the silverview
program
configuration
•
Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
•
Setting Output Processing
The temperature range for the 4 to 20mA output, the averaging period and the peak/
valley hold processing can be set by using the main program menu to select:
[Setup] → [Output Processing]
The temperatures corresponding to 4 and 20mA can then be set (minimum -40°F
(–40°C), maximum 1832°F (1000°C), minimum span 180°F (100°C), maximum span
1832°F (1000°C)).
To minimise the the effects of temperature fluctuations, noise etc on the output
signal, the averaging period can be set between 0 and 60 seconds.
If required, Hold Processing can be applied by setting hold mode to Peak Hold or
Valley Hold and setting a hold period between 0 and 1200 seconds.
•
Setting Target Emissivity
To set an Emissivity value for the SilverU Sensor, use the main program menu to
select:
[Setup] → [Emissivity]
Emissivity can then be set to a value between 0.10 and 1.0, or selected from a pre-
set list of materials by selecting [Data].
7