Cleaning
This product must be disinfected as normal preparation for each patient. Read this entire section before
cleaning the Demi Ultra unit. Failure to follow these cleaning guidelines could cause damage.
Cleaning the LED Light Attachment and Handpiece
Do not autoclave the LED Light Attachment or any part of the curing light. Demi
Ultra parts are not autoclaveable. Never use a blunt instrument to remove composite
from the lens. Doing so will damage the lens. Do not immerse or spray cleaning solution
directly onto components. Instead, spray cleaning solutions on a cloth to wipe. Do not
CAUTION
clean the handpiece when it is resting in the Charging Dock.
Use the Optics Maintenance Kit (part number 21042) to clean and polish the LED Light Attachment lens.
1.
Follow the instructions included with the kit.
Select the cleaning solution. Use a quaternary ammonium compound product like CaviCide
2.
solution or CaviWipes
follow the directions on the package.
Remove the Demi Ultra from the Charging Dock and remove the Light Shield from the Handpiece.
3.
Wet a cloth or paper towel with cleaning solution and wipe onto the Handpiece, LED Light Attachment,
4.
and Light Shield. Avoid the lens at the end of the LED Light Attachment. Do not separate the Handpiece
and the LED Light Attachment.
When cleaning this product, prevent liquids from entering any opening. Wipe carefully so
that liquid does not run, and always clean the Handpiece with the LED Light Attachment
attached to avoid getting moisture or debris in the Magnetic Socket.
CAUTION
Cleaning the Charging Dock and Radiometer
Be sure to disconnect the power cord from the Charging Dock before cleaning. The
charging dock should be disinfected before every charge to prevent cross-contamination.
Do not spray cleaning solution directly onto components. Instead, spray cleaning solutions
on a cloth to wipe.
CAUTION
Disconnect the power cord from the Charging Dock.
1.
Wet a cloth with cleaning solution and wipe solution onto the
2.
Do not wipe or spray the electrical contacts.
3.
Examine the Radiometer Sensor for debris and wipe as necessary
4.
Wipe carefully so that liquid does not run. Do not allow fluid into the port for the electrical cord.
5.
Prevent liquids from entering any opening.
Additional Cleaning Cautions:
Do not immerse any components of this product in solutions.
Use of solutions other than those recommended may damage plastic parts and will void
product warranty.
CAUTION
Do not use: Denatured alcohol, Lysol
solutions.
Do not use cleaning solutions on the electrical contacts.
®
/CaviWipes1 that contain 25% alcohol or less. If using CaviWipes
Dock surfaces .
, phenol, ammonia complex or iodine complex
®
11
/CaviCide1
®
®
/CaviWipes1,