Cassette Removal; Bi-Directional Pumping; Partial Bank Pumping - MANOSTAT CARTER 74-083-00000 Manual De Operación

Sistema de cabeza de bomba de cassette
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

(3) Cassette Removal

Follow this procedure to unload the Cassette.
1. Set Drive to "0" and turn Drive OFF.
2. Push down on latch side of Cassette and pull latch outward to unhook from rod.
3. Pivot Cassette upwards and unhook Hook Leg from rod and remove Cassette.

(4) Bi-Directional Pumping

Fluids can be pumped into and back out of containers by reversing the drive.
WARNING:
Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to
protect operator and equipment.

(5) Partial Bank Pumping

The Pump can be operated with either a partial or
full bank of Cassettes.There is an elastic, VITON
fluoroelastomer O-ring on each of the two metal rods
of the pump (See Figure 7) that holds the Cas-settes
in place on the pump. These O-rings should be
adjusted as follows. If the pump is partially loaded
with Cassettes, slide the O-rings along the rods, up
against the front of the outermost Cassette, to keep
it in place (note position A, for example, in Figure 7).
If the pump is fully loaded with Cassettes, slide the
O-rings to the front of the pump as far as possible
(note position B in Figure 7).
8
®
POSITION A:
(PARTIALLY
LOADED)
O-RING
FIGURE 7 O-RING POSITION
POSITION B:
(FULLY LOADED)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Carter 74-084-00000Carter 72-557-000Carter 72-557-100

Tabla de contenido