G1 - G9
4.5
Generator Shutdown Procedure
Close the external shut off valves.
Press [
] to put the generator into standby.
The menu will change as shown to indicate that the generator is shutting down.
The generator will continue to run for up to 10 minutes in order to allow the cycle to
complete and to ensure that the inlet piping and CMS columns are fully de-pressurised.
Once complete the menu will default to "Standby".
The internal storage receivers of the generator will still be pressurised. These MUST be depressurised if the generator
is to be serviced, shipped or placed into storage.
Warning
Once complete the menu will default to "Standby".
4.5.1 Depressurisation
If the generator is not fitted with the economy option:
With he generator in "Standby" disconnect the generator from the application option.
Observe the "Pressure gauges" located behind the upper service hatch. When all of the pressure gauges read 0 bar the generator is
depressurised.
If the generator is fitted with the economy option:
Disconnect the generator from the application piping and the electrical power supply.
Press and hold [
The generator will begin to depressurise as indicated by the on screen pressure
reading "Po *.*".
When the generator is completely depressurised it will revert to standby mode.
Observe the "Pressure Gauges" located behind the upper service hatch to ensure that
they read 0 bar.
Disconnect the generator from the electrical power supply.
Where applicable disconnect the generator from the compressed air supply.
The compressor heads are very hot. Allow the generator to cool for 30 minutes following shutdown.
Warning
] and reconnect the generator to the electrical power supply.
13
0 0 0 0 0
:
0 0 H r s
S t a n d b y
S h u t t i n g
D o w n
S t a n d b y
P o 0 .
D e p r e s s u r i s i n g
S t a n d b y
R u n
0 0
R u n