ModBUS Register Information
Group Name
Setup/Relay 3/Event
Setup/Relay 3/Event
Setup/Relay 3/Event
Setup/Relay 3/Timer
Setup/Relay 3/Timer
Setup/Relay 3/Timer
Setup/Relay 3/Timer
Setup/Relay 3/Timer
Setup/Relay 3/Timer
Setup/Relay 3/Status
Comm/Net Card
Comm/Net Card
Comm/Net Card
Comm/Net Card
Comm/Net Card
Comm/Net Card
Comm/Net
Card/Addresses
Comm/Net
Card/Addresses
Comm/Net
Card/Addresses
Comm/Net Card/Stats
Comm/Net Card/Stats
Comm/Net Card/Stats
70
Table A-9 Controller ModBUS Registers (continued)
Tag Name
Register # Data Type
On Min Time
40160
Off Max Time
40161
Off Min Time
40162
Hold Type
40163
Sensor Select
40164
Hold Mode
40165
Duration Time
40166
Period Time
40167
Off Delay
40168
Level
40169
Mode
40170
Baud
40171
Stop Bits
40172
Data Order
40173
Min
Response
40174
Time
Max
Response
40175
Time
sc100
40176
Sensor 1
40177
Sensor 2
40178
Good
40179
Messages
Bad
40181
Messages
% Good Mesg
40183
Length
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
2
R/W
Integer
Unsigned
2
R/W
Integer
Float
2
R/W
Description
Sets the min on time (mins)
Sets the max off time (mins)
Sets the min off time (mins)
Sets which Sensor outputs are affected
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being
held/transferred during the timers on time
(this is used when Hold type is set for
single Sensor)
Selects hold outputs vs. set transfer value
during timers on time
Sets the timer on duration time (sec)
Sets the period between timer on events
(mins)
Sets the time the affected Sensor outputs
are held/transferred after the timer turns
off (sec)
Sets the status level which will trigger
the relay
Sets the Modbus mode
(0 = RTU; 1 = ASCII)
Sets the Modbus baud rate (0 = 9600;
1 = 19200; 2 = 38400; 3 = 57600;
4 = 115200)
Sets the number of stop bits (1,2)
Sets the register data order for floats
(0 = Normal; 1 = Reversed)
Sets the minimum response time
(0 to 30 sec)
Sets the maximum response time
(100 to 1000 sec)
Sets the sc100 Modbus Address
Sets the Sensor 1 Modbus Address
Sets the Sensor 2 Modbus Address
Number of good messages
Number of failed messages
% of good messages