Technical Specification - Reer SkinTemp Instrucciones

Termómetro infrarrojo sin contacto 3 en 1
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 13

Technical specification

Measuring technology
Infra-red technology
Measuring range
Body temperature: 32,0°C – 42,9°C
Object/ambient temperature: 0°C – 100°C
Measuring accuracy
Body temperature: +/- 0,2°C (35,5°C - 42,0°C);
+/- 0,3°C (< 35,5°C/ > 42,0°C)
Object temperature: +/- 1°C (15,0°C - 60,0°C);
+/- 2°C (< 15,0°C/ > 60,0°C)
1 second
Measuring duration
Measured value memory
19 measurements
Measured value display
Liquid crystal display (LCD)
30 seconds after last measurement
Power saving mode
Size
133,8 x 32,7 x 23 mm
Weight
39 g (including batteries)
DC 3V, CR2032 (included in scope of delivery)
Batteries
Protection degree
IP22
Operating
15°C – 40°C ambient temperature,
20% - 95% relative humidity,
conditions
70-106 KPa atmospheric pressure
Storing
-25°C – 55°C ambient temperature,
conditions
≤ 95% relative humidity,
70-106 Kpa atmospheric pressure
Metrological control
As a rule, an inspection of the thermometer is not required. For devices used in
medical practice, the German Medical Products Operator Ordinance of 29 June
1998 requires a metrological control after 2 years.
22
Troubleshooting
Problem/symbol
Possible cause
Device cannot be
Batteries are empty.
switched on
Temperature
Sensor is dirty/defect
values wrong or
fluctuating
Low battery level,
measuring is still
possible.
Batteries are empty, it is no
longer possible to measure
using this device.
The operating temperature
is outside the permitted
range of 15°C – 40°C.
Measured body tempera-
ture is over 42.9°C.
Measured object/ambient
temperature is over 100°C.
Measured body tempera-
ture is lower than 32°C.
Measured object/ambient
temperature is lower
than 0°C.
Measures
Insert new batteries.
Check the measuring
sensor and clean it if neces-
sary as described
on page 21.
Replace the
batteries soon.
Replace the batteries
immediately.
Take the thermometer
to a place where the
temperature is inside the
permissible range.
Acclimatise the thermome-
ter for at least one hour.
Check if you are in the
correct measuring mode
(body temperature/object
temperature).
Repeat the measurement.
Ensure the measuring sen-
sor is correctly positioned.
Check if you are in the
correct measuring mode
(body temperature/object
temperature).
Repeat the measurement.
Ensure the measuring sen-
sor is correctly positioned.
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

98020

Tabla de contenido