PROGRAMMING OF THE FUNCTIONING IN pH TYPE
PROGRAMMING AND VISUALIZATION OF ON-OFF FUNCTIONING IN pH TYPE
PROGRAMMING AND VISUALIZATION OF PROPORTIONAL FUNCTIONING IN pH TYPE
Rev 1.0
Time base: 300 sec
Push and quickly release CAL button and scroll the menu using + and – buttons
till the display shows "Function Type". Push CAL to enter the submenu and
then, by pressing + and – button, choice "Function On-Off" and confirm pressing
CAL.
The display will show the value of setpoint stored in the last programming. Modify
it by pressing + and – button and confirm the choice pressing CAL.
The display will show the direction of dosing Acid or Alk stored in the last
programming. Choice the direction by + or – button and confirm pressing CAL.
Select Acid if you want the pump to inject for pH values greater than Setpoint.
Select Alk if you want the pump to inject for values of pH lower than Setpoint.
The display will show the % of maximum flow rate stored in the last programming.
Modify it by pressing + and – button and confirm the choice pressing CAL.
The display will show the value of "Delay On" stored in the last programming.
This is the seconds of delay of motor's start up when the measure will overtake
the setpoint. Modify it by pressing + and – button and confirm the choice pressing
CAL.
Push ESC button to esc the menu and go back in measure with the On-Off
functioning just programmed.
In the On-Off functioning if the pump is dosing the display shows "On" and the %
flow rate programmed on the first row, if the pump doesn't dose the display
shows "Off" and the % flow rate programmed.
On the second row the display writes the measure.
Range of proportionality = 1pH
Time base: 300 sec
Push and quickly release CAL button and scroll the menu using + and – buttons
till the display shows "Function Type". Push CAL to enter the submenu and
then, by pressing + and – button, choice "Function Prop" and confirm pressing
CAL.
The display will show the value of setpoint stored in the last programming. Modify
it by pressing + and – button and confirm the choice pressing CAL.
The display will show the direction of dosing Acid or Alk stored in the last
programming. Choice the direction by + or – button and confirm pressing CAL.
Select Acid if you want the pump to inject for pH values greater than Setpoint.
Select Alk if you want the pump to inject for values of pH lower than Setpoint.
The display will show the value of "Delay On" stored in the last programming.
This is the seconds of delay of motor's start up when the measure will overtake
the setpoint. Modify it by pressing + and – button and confirm the choice pressing
CAL.
Push ESC button to esc the menu and go back in measure with the Proportional
functioning just programmed.