Transcend 3 miniCPAP Quick Guide

Cleaning the Exterior

Follow these instructions to clean the exterior of the Transcend 3 miniCPAP.
1.
Unplug the power supply prior to cleaning and disconnect the device from power cords.
2.
Mix a solution of 5% mild liquid detergent in distilled water (1.6 fl oz liquid detergent
per quart of distilled water). Mild detergent should contain biodegradable anionic
surfactants and no phosphate.
3.
Submerge a lint-free cotton cloth into the detergent solution.
4.
Wring excess water from the cloth then wipe the exterior of the Transcend 3 miniCPAP
device for approximately 20 seconds using a gentle, back and forth wiping motion from
the front to back of the device. Apply firm pressure and ensure contact with all
accessible contact surfaces to adequately remove soil buildup.
5.
Rinse the cloth in clear water to remove residual cleaning solution.
6.
Wring excess water from the cloth then wipe the Transcend 3 miniCPAP using a gentle
front to back wiping motion to remove any detergent solution remaining on its surface.
Wipe the device with a dry, lint-free cotton cloth until the device is fully dry.

Cleaning of Accessories

The following accessory should be cleaned with a 5% solution of mild liquid detergent in
distilled water (1.6 fl oz liquid cleaning detergent per quart of distilled water). Mild
detergent should contain biodegradable anionic surfactants and no phosphate. Follow
these steps to clean the accessories.
Accessory
Air Supply Tube
1.
Fully immerse the air supply tube in the cleaning solution.
2.
While immersed, thoroughly wipe the surface with a lint-free cotton cloth. Apply firm
pressure and ensure contact with all accessible contact surfaces to adequately remove
soil buildup.
3.
Clean the inside of the air supply tube by lifting, then lowering the ends of the tube
while the tube is filled with cleaning solution.
4.
Rinse air supply tube by immersing in distilled water. Move the air supply tube in a
back and forth motion for approximately10 seconds to remove cleaning agent residue.
5.
Rinse the air supply tube in distilled water by fully immersing. Lift, then lower, the
ends of the tube while the tube is filled with water. Repeat this motion for
approximately 10 seconds to remove cleaning agent residue.
6.
Dry the outside of the air supply tube with a dry, lint-free cotton cloth. Allow the tube
to air dry until the inside of the tube is dry. Length of drying time will depend on
ambient conditions.
Periodic Cleaning Cycle
Daily
13
ENGLISH
Product Service Life
3-Month
Page 12