NOUVAG AG
4)
Autoclave packed motor incl. cable at maximum 134° C .
When autoclaving without vacuum, a drying phase must occur. Let motor dry in the bag for at least an
hour at room temperature, with the paper side towards the top.
7.1 Exchanging control unit fuses
Defective control unit fuses can be replaced by the user without outside help. They can be found at the
back of the device in the fuse compartment next to the main switch:
1) Plug out mains cable.
2) Open fuse compartment with a screw driver.
3) Replace defective fuse T 1A.
4) Insert fuseholder again and close fuse compartment.
5) Check displayed mains voltage on fuse compartment.
6) Plug back mains cable.
1. Displayed mains voltage
2. Fuse compartment
3. Clasp fuse compartment
4. Fuse 1
5. Fuse 2
7.2 Discarding instructions
When discarding the device, device components and accessories, please comply with the issued
statutory regulations.
With regard to the preservation of the environment old equipment may be returned to the distributor or
manufacturer.
No. 31739 35/05
7 Maintenance
TCM 3000 BL
7