XT Auto/ English
9. Technical Specifications
I t e m
Power Supply
Pressure Range
Initial Pressure
Maximum Pressure
Minimum Pressure
Ramp Time
Ramp Starting Pressure
Operating Altitude
Dimensions (W x D x H)
Weight
Temperature
Environment
Humidity
Classification:
Note: The manufacturer reserves the right to modify the specification without notice.
Universal power supply, AC100-240V, 50/60 Hz, 0.5-0.3A
4-20 cmH
O (adjustable in 0.5 cmH
2
4-19 cmH
O (adjustable in 0.5 cmH
2
5-20 cmH
O (adjustable in 0.5 cmH
2
4-19 cmH
O (adjustable in 0.5 cmH
2
0-45 minutes (adjustable in 5-minute increment)
4-19 cmH
O (adjustable in 0.5 cmH
2
up to 8,000 ft (2,438 m) when the pressure is set at 4-18 cmH
but limit to 5,000ft (1,524m) when the pressure is set at 18.5 - 20
cmH
O
2
14.5 x 13.0 x 10.0 cm or 5.7" x 5.1" x 3.9"
Approximately 800 g or 1.76 lb
Operating: +5°C to +35°C (+41°F to +95°F)
Storage: -15°C to 50°C (+5°F to +122°F)
Shipping: -15°C to 70°C (+5°F to +158°F)
Operating: 15%RH to 95%RH non-condensing
Storage: 10%RH to 90%RH non-condensing
Shipping: 10%RH to 90%RH non-condensing
Class II
Type BF, Applied Parts: Mask
Not suitable for use in the presence of a flammable anesthetic mixture
IP21: Protected against solid foreign objects of 12.5 mm and greater;
Protected against vertically falling water drops.
Continuous operation.
S p e c i f i c a t i o n s
O increment)
2
O increment)
2
O increment)
2
O increment)
2
O increment)
2
16
O
2