7.3.2-Emissivity
In Object parameter submenu, press "up" and "down" button, highlight "Emiss", press "select" button,
popup emissivity submenu.
"Emiss" sets object emissivity, the value range is 0.01~1.00;
7.3.3-Reflective Temperature
The reflective temperature is important for radiometric temperature measurement. Thermal Imager has
temperature compensation for reflective temperature. To get more accurate temperature measurement, accurately
set the reflective temperature. In most cases, the reflected temperature is identical to the ambient temperature.
Only when objects with strong emissions with much higher temperature are in the proximity of the object being
measured, the reflected temperature must set.
1-In Object parameter submenu, press "up" and "down" button, highlight "Ref".
2-Press "select" button, the reflective temperature will be modified.
3-Press "up" and "down" button to modify the Reflective temperature.
4-Press "Select" to confirm, "Close" to exit without modify.