TROUBLESHOOTING

The compressor must not be repaired except by the Technical
Service of PARI GmbH or a service center expressly author-
ised to do so by PARI GmbH. If the compressor is opened or
manipulated by anyone else, all claims under the warranty
shall be void. In these cases, PARI GmbH will accept no
liability.
Fault
The com-
pressor does
not start up.
No aerosol is
coming out of
the nebuliser.
– 35 –
Possible cause
The power plug is
not plugged into the
mains socket prop-
erly.
The supply voltage is
not suitable for run-
ning the compressor.
The nebuliser nozzle
attachment is
blocked.
The connection
tubing is not connec-
ted properly.
The connection
tubing is leaking.
Remedy
Check whether the power
plug is plugged into the
socket correctly.
Check whether the local sup-
ply voltage matches the
voltage data marked on the
compressor identification
label.
Clean the nebuliser.
Check that the tubing con-
nectors are connected firmly
to the compressor and the
nebuliser.
Replace the connection
tubing.