MP6025647-2AF1 / A
Immunity Test
Electrostatic dis-
charge (ESD)
IEC
61000-4-2
Electrical fast
transient/burst
IEC
61000-4-4
Surge
IEC
61000-4-5
Voltage dips,
short interrup-
tions and volt-
age variations
on power supply
lines
IEC
61000-4-11
Power frequen-
cy (50/60 Hz)
magnetic field
IEC
61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 Test
Compliance
Level
±6 kV indirect
±6kV indirect
(30-60% relative
humidity)
±30 kV air
±8 kV air
(<5% relative
humidity)
±2kV for power
Not applicable
supply lines
±1 kV for in-
put/output lines
Not applicable
±1 kV lines(s) to
Not applicable
line(s)
±2 kV line(s)
Not applicable
to earth
<5% U
(>95%
Not applicable
T
dip in U
) for 0.5
T
cycle
40% U
(60%
T
Not applicable
dip in U
) for 5
T
cycles
70% U
(30%
T
dip in U
) for 25
Not applicable
T
cycles
<5% U
T
Not applicable
(>95% dip in U
)
T
for 5 seconds
3 A/m
3 A/m
Electromagnetic Environment -
Level
The MiniLink transmitter should not
be affected by electrostatic dis-
charge that might occur under nor-
mal conditions of use.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
-14-
Guidance