3 Press the circuit breaker to the desired position: ON ( | ) or OFF (O).
Figure 4
The dock circuit breaker is on the side toward the wall.
To reset the stand module circuit breakers
One or both of the circuit breakers on the stand module trip if there is a power event (for example, a power
surge) and may need to be reset.
1 Undock the stand, and remove the ultrasound system and any loose items.
2 Lock the wheels, and tip the stand onto its side, allowing access to its underside.
WARNING
3 Press in on each circuit breaker switch until it closes. See
On a correctly closed circuit breaker, the switch is pressed in.
On a tripped circuit breaker, the switch is popped up and exposes a white plastic area.
Note
6
For SonoSite X-Porte, we recommend keeping the system upright and
accessing the circuit breakers from underneath. Tipping the system could
cause injury.
If you cannot press down the switch and if no white area is exposed, the circuit
breaker is closed.