SPORT-ELEC FREE ACTION Manual Del Usuario página 56

Ocultar thumbs Ver también para FREE ACTION:
Tabla de contenido

Publicidad

GUIDANCE AND MANUFACTURER'S DECLARATION
ELECTROMAGNETIC EMISSIONS
The Free Action is intended for use in the electromagnetic environment specified below. The customer or the user of
the Free Action should assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
The Free Action uses RF energy only for its internal function. Therefore, its RF
RF emissions
Group 1
emissions are very low and are not likely to cause any interference in nearby
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Conform
The Free Action is suitable for use in all establishments, including domestic
IEC 61000-3-2
Class A
establishments and those directly connected to the public low-voltage power
Voltage fluctuations/ flicker
emissions
Conform
IEC 61000-3-3
GUIDANCE AND MANUFACTURER'S DECLARATION
ELECTROMAGNETIC IMMUNITY
The Free Action is intended for use in the electromagnetic environment specified below. The customer or the user of
the Free Action should assure that it is used in such an environment.
IMMUNITY
IEC 60601 TEST LEVEL
COMPLIANCE LEVEL
TEST
Electrostatic
±6 kV contact
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
±2 kV for power supply lines
±2 kV for power supply
and dc lines
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable for input/
±1 kV for input/output lines
Surge
±1 kV line(s) and neutral
±1 kV line(s) and neutral
IEC 61000-4-5
N/A(§7.2 EN61000-4-5)
± 2 kV line(s) to earth
<5 % U
(>95 % dip in U
)
<5 % U
T
T
for 0,5 cycle
Voltage
dips, short
40 % U
(60 % dip in U
)
40 % U
T
T
interruptions
for 5 cycles
and voltage
variations
70 % U
(30 % dip in U
)
70 % U
on power supply
T
T
for 25 cycles
input lines
IEC 61000-4-11
<5 % U
(>95 % dip in U
)
<5 % U
T
T
for 5s
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE - U
is the a.c. mains voltage prior to application of the test level
T
110
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
electronic equipment.
supply network that supplies buildings used for domestic purposes.
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
±8 kV air
relative humidity should be at least 30 %.
lines
Mains power quality should be that of a typical
output lines
commercial or hospital environment.
(>95 % dip in U
)
T
T
for 0,5 cycle
Mains power quality should be that of a typical
(60 % dip in U
)
T
T
commercial or hospital environment. If a dips or an
for 5 cycles
interruption of mains power occurs, the current of
the Free Action may be dropped off from normal
(30 % dip in U
)
T
T
level, it may be necessary to use uninterruptible
for 25 cycles
power supply or a battery.
(>95 % dip in U
)
T
T
for 5s
Power frequency magnetic fields should be at
3A/m
levels characteristic of a typical location in a typical
commercial or hospital environment.
GUIDANCE AND MANUFACTURER'S DECLARATION
ELECTROMAGNETIC IMMUNITY
The Free Action is intended for use in the electromagnetic environment specified below. The customer or the user of
the Free Action should assure that it is used in such an environment.
IMMUNITY
IEC 60601
COMPLIANCE
TEST
TEST LEVEL
LEVEL
Conducted RF IEC
3 Vrms
3V
61000-4-6
150 kHz to 80 MHz
Radiated RF IEC
3 V/m
3 V/m
61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Free Action is used exceeds
the applicable RF compliance level above, the Free Action should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the Free Action.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
RECOMMENDED SEPARATION DISTANCES bETWEEN PORTAbLE AND MObILE RF
COMMUNICATIONS EqUIPMENT AND ThE FREE ACTION
The Free Action is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Free Action can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Free Action as recommended below,
according to the maximum output power of the communications equipment.
SEPARATION DISTANCE ACCORDING TO FREqUENCY OF TRANSMITTER (M)
RATED MAXIMUM OUTPUT
POWER OF TRANSMITTER
150 KhZ TO 80 MhZ
80 MhZ TO 800 MhZ
(W)
d = 1.17
0.01
0.12
0.1
0.37
1
1.17
10
3.69
100
11.67
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 - At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
Portable and mobile RF communications equipment should be used no
closer to any part of the Free Action, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 1.17
d = 1.17
80 MHz to 800 MHz
d = 2.33
80 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation Distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,a should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment marked with
the following symbol:
800 MhZ TO 2.5 GhZ
d = 1.17
d = 2.33
0.12
0.23
0.37
0.74
1.17
2.33
3.69
7.38
11.67
23.33
111

Publicidad

Tabla de contenido
loading

Tabla de contenido