THE ADAPTOR PAD and ROHO DRY FLOATATION Special Application Cushioning Devices Operation Manual
Recliner System
Intended Use: The ROHO
support surface that utilizes DRY FLOATATION
seated shape to provide localized skin/soft tissue protection and an environment to facilitate wound healing.
There is no weight limit, yet the cushion must be properly sized to the individual.
Instructions:
1. Place the Recliner System on the chair with the air cells
facing up and the non-skid surface of the cover facing the
chair seat and chair back.
When possible, use the hook or loop fasteners to
further secure the cover to the seating surface. On the
back cushion, hook and loop fasteners are located inside the
cover and on the back of the cushion.
2. Slide the hand pump nozzle over the inflation valve. Open
the inflation valve, at least one full rotation. Inflate until all of
the air cells feel firm. Close the inflation valve until it comes
to a stop. Remove the hand pump. Make sure the inflation
valve is closed. Repeat for all inflation valves.
3. Confirm that the individual is in their usual position.
Confirm correct placement of the Recliner System.
4. Slide your hand between the cushion and the individual
and feel for bony prominences. With your hand in place,
open the inflation valve, releasing air as prescribed. Close
the inflation valve. Repeat for all inflation valves.
Note: The inflation valve will spin freely when open.
CHECK INFLATION! Make sure that there is at least ½ in.
(1.5 cm) of air between the cushion base and the individual's bony
prominences. Follow the prescriber's recommended inflation
levels.
Recliner Cushion Cover Removal: Unsnap and remove the covers from the cushions.
Recliner Cushion Cover Replacement:
1. The zipper on the Back Cushion cover should be closed. For the Seat Cushion and Back
Cushion: The non-skid cover material should face up and the snapped opening of the cover
should face you. The air cells should face down: Insert each cushion into its cover, inflation
valve(s) first. Pull the inflation valve(s) and Yellow Rope through the corner opening(s).
The inflation valves will protrude at the front of the Seat Cushion and at the top of the Back
Cushion.
2. Place the two cushions with the snapped openings next to each other. Snap the Back
Cushion to the Seat Cushion. Snap the bottoms of the covers together. Turn the cushions
over so the air cells face up. Snap the tops of the covers together. The cushions should lie
flat inside the covers.
Recliner System (Recliner System) is an adjustable, air-filled, cellular-design,
®
technology and is intended to conform to an individual's
®
7
Shown
without
cushion
covers.
The seated figure shows
standard orientation.
Open
Close