Mouth specula
Not all mouth specula with a glass mirror in them are suitable
for machine reprocessing.
Follow the manufacturer's instructions.
Rhodium-coated mouth specula, because of their delicate surface,
must be loaded in such a way that the mirror surfaces cannot sus-
tain mechanical damage during reprocessing, e.g. by knocking
against other instruments.
Place the mouth specula individually into the 13 x 13 mm (E 337/1)
or 14 x 20 mm (E 802/1) compartments (one speculum in each
compartment).
E 337/1 insert and A 830 insert
The A 830 insert (optional accessory) must be used when repro-
cessing separable mouth specula. The insert is supplied with its own
set of operating instructions.
Unscrew the work ends with the mirror surfaces before repro-
cessing.
en - Areas of application