POLYLUX fiber optic light handpiece

8.1

Care and cleaning of the handle sleeve

1
8.2

Care and cleaning of the light guide

59 06 511 D 3386
D 3386.103.01.06.02
8 POLYLUX fiber optic light handpiece
8.1 Care and cleaning of the handle sleeve
The handle sleeve with the fiber optic rod
can be unscrewed (counterclockwise) from the
handpiece for cleaning and sterilizing.
Sterilize only in an autoclave at 135°C, 2.1 bar.
ATTENTION
NEVER exceed 150°C! DO NOT subject the handle
sleeve to ultrasonic cleaning or chemical vapor
sterilization!
Prior to sterilization
clean the light exit surface (1) with alcohol (50%).
ATTENTION
DO NOT place the handle sleeve in disinfectant
solutions!
The light guide (handpiece hose) can be pulled out for
cleaning and disinfecting from the lamp module under
the dentist or assistant element.
ATTENTION
Always grasp the gray anti-kink sleeve (2) of the light
guide, and NEVER the hose!
Do not squeeze, twist or kink the handpiece hose!
Do not exceed a minimum bending radius of
80 mm.
Clean
with alcohol (50%) or with an aqueous detergent
solution.
Clean the light entry and exit surfaces (3) with alcohol
(50%).
Disinfection by wiping
is acceptable.
ATTENTION
NEVER sterilize or thermodisinfect the light guide!
DO NOT place the light guide in a liquid bath or subject it
to ultrasonic cleaning!