10 Checking the safety switches
9.3 Care and cleaning of the monitor
10
Checking the safety switches
28
To ensure the safety of the practice team and the
patients, please check the safety switches installed in
the unit every 6 months.
Move the chair or the backrest down for this check.
• Raise the leg support:
The chair movement stops, then there is a corrective
movement upwards for around 3sec.
• Raise the backrest:
The backrest movement stops, then there is a correc-
tive movement upwards for around 3sec.
• Raise the motor-adjustable headrest:
The backrest movement stops, then there is a correc-
tive movement upwards for around 3sec.
• Swing in the swiveling cuspidor (optional):
The chair movement stops.
ATTENTION
If one of the safety switches does not stop the chair
movement or the backrest movement, the treatment
center must not be used for dental work!
Please contact your service engineer.